Model | MCO-230AIC |
External Dimensions(W x D x H) | 770 x 730 x 905 mm |
Internal Dimensions(W x D x H) | 643 x 523 x 700 mm |
Volume | 230 liters |
Net Weight | 90 kg |
CO2 Control Range & Fluctuation | 0% ~ 20%, ±0.15% |
Display | LCD Touch Screen |
Shelves | 4 |
External Dimensions (W x D x H) | 770 x 730 x 905 mm |
Internal Dimensions (W x D x H) | 643 x 523 x 700 mm |
Volume | 230 liters |
Net Weight | 90 kg |
Temperature Control Range & Fluctuation | AT +5°C ~ +50, ±0.1 |
Temperature Uniformity | ±0.25°C |
CO2 Control Range & Fluctuation | 0% ~ 20%, ±0.15% |
O2 Control Range & Fluctuation | - |
Humidity Level & Fluctuation | 95% RH, ±5 |
Temperature Sensor | H2O2 Decontamination |
Sterilisation Method | Thermistor |
CO2 Sensor | Dual IR |
O2 Sensor | - |
Display | LCD Touch Screen |
Exterior Material | Painted Steel (rear cover not painted) |
Interior Material | Stainless Steel Copper-Enriched Alloy |
Insulation Material | Extruded polystyrene |
Heating Method | Direct Heat & Air Jacket System |
Outer Doors | 1 |
Outer Door Lock | Optional |
Field Reversible Door | Included |
Inner Doors | 1 gastight - made of tempered glass |
Shelves | 4 x Stainless Steel Copper-enriched Alloy |
Shelf Dimensions (W x D x H) | 620 x 450 x 12 mm |
Max. Load - Per Shelf | 7 kg |
Max. Shelf Capacity | 10 |
Access Port | 2 |
Access Port Position | Rear |
Access Port Diameter | 30 |
Power Failure | R |
Out of Temperature Setting | V-B-R |
High Temperature | V-B-R |
Out of CO2 Setting | V-B-R |
Out of O2 Setting | - |
Door Open | V-B |
Power Supply | 230 V |
Frequency | 50 Hz |
Noise Level | 25 dB |